Home » Dental Equipment Supplies » Autoclaves Sterilizers » Large Laboratory Autoclave Sterilizers » Large Sterilizers (Compact Series) - 310 Liters

Large Sterilizers (Compact Series) - 310 Liters

Item# 02155TUT

Sterilization in a laboratory environment has its unique requirements.  The types of loads that need to be sterilized are varied and include liquids, glassware, instruments, porous loads and additional laboratory items.  Choosing the right steam sterilizer depends on several considerations including: the type of loads, frequency of use, available services and load volume.

  • Chamber Dimensions (508W x 508H x 1210D)  in mm
  • Chamber Volume in Liters (310)

Features:

  • The Compact autoclave series is designed to cover a large range of applications for laboratories, research institutes and biotechnology facilities. 
  • The autoclave operates with saturated steam as the sterilizing agent with a temperature range from  105°C   (221°F)   to   137°C   (279°F) and a working pressure that meets AMSE and PED requirements.
  • Each autoclave model is available with either single or double door.

The Compact series is available with two door options:

  • Fully automatic vertical sliding door
  • Manual Hinged door
  • The laboratory autoclaves are designed to meet BL3 and BL4 bio-safety levels. 
  • Bio-Hazard system allows for an effluent decontamination cycle that sterilizes contaminated laboratory waste.

Engineered with People in Mind

  • Our quality features enable the convenience and durability needed to operate an autoclave with complete peace of mind.
  • The fully jacketed chamber is constructed of long lasting 316L grade stainless steel with superior corrosion resistance.
  • In addition, the generator and piping are constructed of stainless steel.
  • The pneumatic valves are air pressure operated, reducing maintenance.
  • They are safer and more reliable, eliminating the requirement for high voltage.
  • If no buttons or switches are operated for four hours, the autoclave will switch to standby mode.

Door Safety:

  • The doors are designed with independent mechanical and digital safety features that guarantee a safe working of the sterilizer.
  • A safety device prevents the operator from opening the door when chamber is pressurized
  • Steam is not allowed into the chamber when the door is open
  • A cycle will not start if the door is open or not properly closed
  • The door cannot open until chamber pressure reaches room pressure
  • Sliding Door Safety -The sliding door progress will automatically stop if an obstruction is detected
  • Double Door Safety – interlocks prevent both doors from being opened simultaneously  
  • General Safety Features:
  • Double Independent Monitoring: The combined digital and mechanical monitoring provides a cross reference and guarantees accurate results.
  • The operator has two independent means to monitor temperature and pressure.
  • Safety Valves: Both the chamber and the jacket are equipped with safety valves –if the pressure exceeds the allowed limit the safety valves will discharge.
  • Built-in Steam Generator Safety: A water level monitoring system maintains a constant water level and ensures safe operation of the heaters.
  • Emergency Shut-Off: Easily accessible emergency switches for immediate and safe shut down of the sterilizer. 
     

ADMC Software

  • The ADMC software allows complete control and monitoring of up to eight sterilizers.
  • The software retrieves data, creates graphs, tables and print outs, controls the autoclave and shows real-time data.

Barcode Reader

  • The barcode reader allows material identification.
  • The data is transferred to the ADMC and linked to the sterilization cycle parameters for streamlined instrument handling.

Chart Recorder

  • The high speed chart recorder is an independent temperature and pressure measuring system.
  • It features analog/mechanical trend records and print-outs in color. Simple operation with an easy-view display allows one to key-in various items of set data.

Data Logger

  • The data logger is the hospital’s protection, making documentation even more convenient and reliable.
  • It stores over 100,000 hours of cycle data, displayed either graphically or numerically.

Touch Screen

  • The touch screen in combination with ADMC-software displays cycle data in both a textual or graphical interface and provides a continuous visual guide of the progress of each stage in the sterilization cycle.
  • Provides a visual display of all piping for monitoring purposes.

Liquid Loads

  • The laboratory sterilizers are designed for treating sensitve liquid loads.
  • The autoclaves ensure that the correct sterilization temperature is attained when sterilization starts and that sterilization is initiated only when a temperature probe placed inside the medium reaches the preset threshold temperature.
Large Sterilizers (Compact Series) - 310 Liters-02155TUT
Your Price: Call For Pricing
1-877-706-4480
Unit: Single

Product in stock
QTY   Add to Cart